Spectrolab Tapping points EM55-1...4


Product Features
- Wall-mounting tapping points / outlet points
- For non-corrosive gases up to quality 6.0
- Laboratory-style design
- Ergonomically designed
- Filter at the process gas inlet valve
- Diaphragm-type shut-off valves optimised for low internal volume with On/Off position indicator
- Suitable for ECD-applications
- Regulator with high control accuracy
- Designed for easy installation
- Approved for use with oxygen
- Tapping point can be extended into multiple tapping points
- Top-inlet or bottom-inlet configurations available
Technical Data
Inlet pressure P₁
max. 200 bar
Outlet pressure P₂
1,5/4/10/20/50 bar
Material
Body regulator and valve: chrome-plated brass
Valve seat regulator: PA
Valve seat shut-off valve: PVDF
Diaphragm regulator: Hastelloy C276
Diaphragm valve: Hastelloy C276
Soft goods: FKM
Filter: Sintered SS 316L
Valve seat regulator: PA
Valve seat shut-off valve: PVDF
Diaphragm regulator: Hastelloy C276
Diaphragm valve: Hastelloy C276
Soft goods: FKM
Filter: Sintered SS 316L
Ports
1/4”-NPT female
Temperature range
-20°C to +60°C
Leak rate
(to atmosphere) 1 x 10⁻⁸ mbar l/s He
(via seat) 1 x 10⁻⁶ mbar l/s He
(via seat) 1 x 10⁻⁶ mbar l/s He
Pressure gauges
Safety pressure gauge ISO5171/cl1.6/NG50
Weight
2,5 kg
Downloads
HTML