Spectrolab Point of use EM55-1...4

Product Features

  • Wall-mounting Point of use
  • For non-corrosive gases up to quality 6.0
  • Laboratory-style design
  • Ergonomically designed
  • Filter at the process gas inlet valve
  • Diaphragm-type shut-off valves optimised for low internal volume with On/Off position indicator
  • Suitable for ECD-applications
  • Regulator with high control accuracy
  • Designed for easy installation
  • Approved for use with oxygen
  • Point of use can be extended into multiple tapping points
  • Top-inlet or bottom-inlet configurations available

Gas supply equipment for non-corrosive gases and gas mixtures up to grade 6.0

    Technical Data
    Inlet pressure P₁
    max. 2900 psi
    Outlet pressure P₂
    22/60/145/290/725 psi
    Material
    Body regulator and valve: chrome-plated brass
    Valve seat regulator: PA
    Valve seat shut-off valve: PVDF
    Diaphragm regulator: Hastelloy C276
    Diaphragm valve: Hastelloy C276
    Soft goods: FKM
    Filter: Sintered SS 316L
    Ports
    1/4”-NPT female
    Temperature range
    -4°F to +140°F
    Leak rate
    (to atmosphere) 1 x 10⁻⁸ mbar l/s He
    (via seat) 1 x 10⁻⁶ mbar l/s He
    Pressure gauges
    Safety pressure gauge ISO5171/cl1.6/NG50
    Weight
    5.5 lb
    CRN
    0C22630