Spectrolab Compact laboratory tapping points EM15


Product Features
- The special laboratory tapping points incorporate the functions shut-off, pressure regulation and pressure indication in one compact, ergonomic unit.
- EM15: for non-corrosive gases up to quality 6.0
EE15: for corrosive gases and gas mixtures with corrosive components up to quality 6.0 - The pressure regulator is diaphragm sensed for outlet pressures up to 10 bar and piston sensed for higher outlet pressure values.
- The acetone resistant pressure gauge is safely integrated into the adjusting handwheel to create an extremely compact device.
- Integrated shut-off valve in the rear-wall connector allows the preparation of the tapping point without pressure regulator.
- Quick and easy mounting or disassembly of the regulating unit with filled gas piping.
- Diaphragm shut-off valve with position indicator
- Optional flow control / shut-off valve in the outlet
- All gas-wetted components have undergone the special SPECTROCLEAN© cleaning process and have been thoroughly baked out.
- For ECD-applications the devices can be treated in an extended cleaning process.
- All equipment has been 100%-helium-leak-tested using a mass-spectrometer.
- All components are plastic-covered resistant to acid and alkaline solutions.
Technical Data
Material
Body: brass
Diaphragms: Hastelloy C276
Other gas-wetted surfaces: brass / SS 316L (1.4404)
Valve cone: SS 316L (1.4404)
Valve seat: PTFE / PVDF
Body coating: Polypropylene GB30
Diaphragms: Hastelloy C276
Other gas-wetted surfaces: brass / SS 316L (1.4404)
Valve cone: SS 316L (1.4404)
Valve seat: PTFE / PVDF
Body coating: Polypropylene GB30
Temperature range
-30°C to +60°C
Leak rate
1 x 10-8 mbar l/s He
Pressure ranges
Inlet pressure P1
P2 to 10 bar: max. 40 bar
P2 > 10 bar: max. 100 bar
Max. outlet pressure P2: 1,0 / 1,5 / 2,5 / 5 / 10 / 16 / 25 / 65 bar
P2 to 10 bar: max. 40 bar
P2 > 10 bar: max. 100 bar
Max. outlet pressure P2: 1,0 / 1,5 / 2,5 / 5 / 10 / 16 / 25 / 65 bar
Downloads
Data Sheets