Laboratory-point of use EE15


Product Features
- The special laboratory point of use incorporate the functions shut-off, pressure regulation and pressure indication in one compact, ergonomic unit.
- EE15 for corrosive gases and gas mixtures with corrosive components up to quality 6.0
- The pressure regulator is diaphragm sensed for outlet pressures up to 145 psi and piston sensed for higher outlet pressure values.
- The acetone resistant pressure gauge is safely integrated into the adjusting handwheel to create an extremly compact device.
- Integrated shut-off valve in the rear-wall connector allows the preparation of the tapping point without pressure regulator.
- Quick and easy mounting or disassembly of the regulating unit with filled gas piping.
- Diaphragm shut-off valve with position indicator
- Optional flow control / shut-off valve in the outlet
- All gas-wetted components have undergone the special SPECTROCLEAN© cleaning process and have been thoroughly baked out.
- For ECD-applications the devices can be treated in an extended cleaning process.
- All equipment has been 100%-helium-leak-tested using a mass-spectrometer.
- All components are plastic-coated resistant to acid and alkaline solutions.
Technical Data
Materials
Body: stainless steel 316L
Diaphragms: Hastelloy C276
Other gas-wetted surfaces: stainless steel 316L
Valve stem: stainless steel 316L
Valve seat: PTFE
Body coating: Polypropylene GB30
Diaphragms: Hastelloy C276
Other gas-wetted surfaces: stainless steel 316L
Valve stem: stainless steel 316L
Valve seat: PTFE
Body coating: Polypropylene GB30
Temperature range
-4°F to +140°F
Leak rate
1 x 10-8 mbar l/s He
Pressure ranges
Inlet pressure P₁
P₂ up to 145 psi: max. 580 psi
P₂ > 145 psi: max. 870 psi
Max. outlet pressure P₂: 15 / 22 / 36 / 73 / 145 / 230 / 360 / 950 psi
P₂ up to 145 psi: max. 580 psi
P₂ > 145 psi: max. 870 psi
Max. outlet pressure P₂: 15 / 22 / 36 / 73 / 145 / 230 / 360 / 950 psi